Short Description:
Bedside monitor for hospital with 5 parameters
Application Range:
Adult/Pediatric/Neonatal/Medicine/Surgery/Operating Room/ICU/CCU
Display: 12.1 inch TFT sreen
Parameter:Spo2, Pr, Nibp, ECG, Resp, Temp
Optional: Etco2, Nellcor Spo2, Suntch Nibp, Touich Screen, Wifi function, Recorder, Trolley, Wall Mount
Language: English, Spanish, Portugal, Poland, Russian, Turkish, French, Italian
Display
Size : 12.1 inch
Number of waveforms Up to 12
Display layout and colors User-configurable
Controls Trim Knob /Touch optional
Parameters and modules
Parameters Modules2
ECG
Resp
SpO2
NIBP
Temp
2 channel InvBP/optional
Sidestream CO2 /Mainstream Optional
Entropy E-Entropy4
ECG
Leads available 3-lead configuration: I, II, III
5-lead configuration: I, II, III, aVR,
aVL, aVF and V
Sweep speed 12.5, 25 or 50 mm/s
Gain range 0.5x, 1x, 2x and 4x
Heart rate accuracy 20 to 300 bpm, ±3% or ±3 bpm,
whichever is greater
Bandwidth
ECG filter Monitor: 0.5 to 40 Hz
ST: 0.05 to 40 Hz
Diagnostic: 0.05 to 145 Hz
Moderate: 0.5-20 Hz
Pacemaker detection Voltage range: 2 to 700 mV
Pulse width: 0.5 to 2 ms
Arrhythmia Alarms
Lethal Alarms Asystole, V Fib/ V Tach, V Tach
HR Alarms Brady, Tachy
Ventricular Alarms VT>2, R on T, V Brady, Couplet,
Bigeminy, Accelerated Ventricular,
Trigeminy, Multifocal PVCs
Atrial Alarms A Fib, Missing beat, Pause, Irregular,
ST segment analysis
Numeric range -9 to +9 mm (-0.9 to +0.9 mV)
Accuracy ±0.2 mm or ±10%, whichever is
greater, within the measurement
range of -8 to 8 mm
Numeric resolution 0.1 mm (0.01 mV)
Impedance respiration
Range Adult/pediatric: 4 to 120 breaths/
min
Neonate: 4 to 180 breaths/min
Accuracy ±5% or ±5 breaths/min, whichever
is greater
Gain range 0.1 to 5 cm/Ohm
SpO2
Digital /Analog SpO2
Measurement range
Pulse oximetry 1 to 100%
Pulse rate 30 to 250 bpm
PI (Perfusion Index) 0 to 32
Measurement accuracy
Saturation Without motion-adult/pediatric
Finger sensor: 70 to 100% ±2%
Without motion-neonate:
70 to 100% ±3%
With motion-adult/pediatric/
neonate: 70 to 100% ±3%
Low perfusion-adult/pediatric:
70 to 100% ±3% (<70% unspecified)
Pulse Rate Without motion: ±2 bpm
(Adult/Pediatric/Neonatal)
Nellcor OxiMax /optional
Measurement range
Pulse oximetry 1 to 100%
Pulse rate 20 to 250 bpm
Measurement accuracy
Saturation Adult: 70 to 100% ±2%
Neonate: 70 to 100% ±3%
Low perfusion: 70 to 100% ±2%
<70% unspecified
Pulse Rate ±3 bpm