TECHNICAL SPECIFICATIONS:
Quality Standards and Classification
Anti-electroshock degree:
ClassⅠequipment
(internal power supply)
TEMP/SpO 2 /NIBP: BF
ECG/Resp: CF
Application Range:
Adult/Pediatric/Neonatal/Medicin
e/Surgery/
Operating Room/ICU/CCU
Display:
14.0inch TFT LCD screen
Resolution: 1366*768
Display: 14.0inch TFT LCD screen
Resolution: 1366*768
SPO2 :
Measuring range 0~100%
Resolution 1%
The accuracy is within the range of 70% to 100%, and
the measurement error should be ±2%;
In the range of 0% to 69%, the measurement error is
ECG:
Gain: X0.25, X0.5, X1, X2, Auto
Waveform speed: 12.5mm/s, 25mm/s,
50mm/s
Sensitivity>200uV (peak-to-peak)
bandwidth:
Diagnostic mode: 0.05~130Hz
Monitoring mode: 0.5~40Hz
Operation mode: 1~20Hz
Common Mode Rejection Ratio:
Diagnostic mode: >90dB
Monitoring mode: >100dB
Operation mode: >100dB
Electrode polarization voltage: ±300mV
Input impedance≥5M
DC bias voltage: ±300mV
Calibration signal: 1mV (peak-to-peak),
accuracy ±5%
ST segment detection measurement:
Measurement range: -2.0mV~+2.0mV
Measurement accuracy: within the range of
-0.8mV~+0.8mV, the measurement error is
±0.02mV or ±10%, whichever is greater,
other ranges are undefined.
ST Segment:
Maximum upper limit 2.0mv
Minimum lower limit -2.0mv
Single adjustment amount 0.1mv
Pulse Rate :
Measuring range 30~300bpm
Resolution 1bpm
Accuracy ±3bpm
Update time 1s
Heart Rate:
Adult 15 to 300bpm (beats/minute)
Neonatal/Child 15 to 350 bpm (beats/minute)
Accuracy ±1% or ±1bpm, whichever is greater
Adult/Pediatric/Neonatal/Medicin
e/Surgery/
Operating Room/ICU/CCU
Electrical Specifications:
Power supply voltage: AC 100~240V
AC, power frequency: 50/60Hz,
maximum input power 70VA
Fuse is F2AL 250V
Working method: continuous work
Built-in lithium battery: d.c11.1V not
less than 2000mA, working time not
less than 2 hours.
Physical Dimensions:
Product Size 336*262*99 mm
product weight 2.6kg
School limit list : 400 groups
Waveform playback : 6000 seconds
Trend table : 7 days
Trend graph: up to 7 days
Alarm review: 60 groups
Temp:
Range 0~50C
Resolution 0.1C
Accuracy ± 0.1C (excluding sensor error)
Accessories:
NIBP cuff & tube
ECG cable & electrodes
SpO 2 sensor
TEMP probe
Lithium-ion battery
Power cable
Operator’s manual
End-tidal CO2 (optional)
Recorder (optional)
Dual IBP, Nellcor (optional)
Respiration rate
Temperature
Measuring Range and Precision
Range
NIBP:
Adult Systolic blood pressure 40~270mmHg
Diastolic blood pressure 10~215mmHg
Mean blood pressure 20~235mmHg
Static pressure range 0~300mmHg
Static pressure precision ±3mmHg
Pressure precision:
Max. average error: ±5mmHg;
Max. standard deviation: 8mmHg